Side-lobe free 4Pi-Microscopy
- High Resolution Optical Microscopy Group, Deutsches Krebsforschungszentrum
marion [dot] lang [at] dkfz-heidelberg [dot] de
In 4Pi-microscopy, the wavefronts of two opposing high aperture objective lenses are coherently added in the focal plane and/or at a common detector point. The net result is a 3-7 times sharper main focal maximum representing a similar improvement of the axial resolution of far-field fluorescence microscopy. However, the main maximum is also accompanied by axial sidelobes of 25-50 % height that lead to imaging artefacts, unless the image data is mathematically postprocessed for sidelobe removal. Therefore 4Pi-imaging has always required the mathematical postprocessing of its raw image data. Here we will discuss a new implementation of 4Pi-microscopy that synergistically combines high aperture angle lenses with a new embedding medium. The observed side-lobes are < 10% of the main maximum, meaning that deconvolution is basically no longer required. As a side-effect, phase adjustment is greatly simplified due to the strongly peaking intensity at the preferred constructive-constructive 4Pi mode of type C.
@inproceedings{lang2006p48,
author = {Lang, Marion and Engelhardt, Johann and Hell, Stefan},
title = {{Side-lobe free 4Pi-Microscopy}},
booktitle = {DGaO-Proceedings 2006},
series = {DGaO-Proceedings},
eventtitle = {107. Jahrestagung der DGaO},
eventdate = {2006-06-07/2006-06-10},
venue = {Weingarten},
publisher = {Deutsche Gesellschaft für angewandte Optik e.V.},
year = {2006},
pages = {P48},
url = {https://dgao-proceedings.de/paper/107-p48}
}